Rotograph Evo D Product Data – Rev. 6 Page 1 of 11
Product Data
Digital Panoramic / Cephalometric X- Ray unit
General Features
Movement technology: multi-motor with digital trajectory control
Patient alignment: through two laser pointers that allow to locate the
reference planes: mid-sagittal, Frankfurt
Standard Examination Programs
Adult panoramic
Child panoramic exposure parameters reduction
TMJ open/close mouth: 4 slices are taken in the same image: left/right condyle,
open/close mouth. Condyles are examined in lateral projection
Maxillary Sinus P-A: one P-A projection, where both the maxillary sinuses are repre-
sented.
In every program the compensation of spinal column is obtained by means parameters modu-
lation, optimized in function of the anatomic program.
Optional “Evo XP eXtended Programs”
Adult half-panoramic (right and left)
Child half-panoramic (right and left)
Improved orthogonality dentition: panoramic projection limited to the dentition obtained
with X-ray beam constantly perpendicular to the arch. It allows to reduce superimposi-
tion of adjacent teeth and to improve visualization of possible interproximal caries.
Frontal dentition: panoramic limited to the frontal dentition (canine to canine), that allows
to improve the detail definition on incisors.
Low dose panoramic: panoramic with reduced angle of rotation to exclude the ascending
ramus from the image. The result is a panoramic limited to the dentition area using a
reduced patient dose.
Bite-wing (right, left, double): the right or left projection allows the examination of
lateral dentition (from eighth to fourth approximately), with optimized trajectory of ro-
tating arm for a higher orthogonality of the x-ray beam on the adjacent teeth, to im-
prove visualization of possible interproximal caries. Double Bite-wing projection per-
forms both Bite-wing views in sequence, joining them on the same image.
Optional “IMPLANT “ Package – Linear tomography for implantology
The exam procedure consists of 3 transversal exposures
Transversal layer thickness: 4 mm (for incisors and canines)
5 mm (for premolars and molars)
Distance between transversal layers: 4 mm (for incisors and canines)
6 mm (for premolars and molars)
Note: the exam is performed through a specific positioner maintaining the examined zone at
the center of useful area.
Page view 0
1 2 3 4 5 6 ... 10 11

Summary of Contents

Page 1

Rotograph Evo D Product Data – Rev. 6 Page 1 of 11 Product Data Digital Panoramic / Cephalometric X- Ray unit General Features Movement t

Page 2 - Image Magnification

Rotograph Evo D Product Data – Rev. 6 Page 10 of 11 Dimensions Pan version (all quotes in mm and inches) 1266 (49,83")÷1600 (63,00")

Page 3

Rotograph Evo D Product Data – Rev. 6 Page 11 of 11 Dimensions Ceph version (all quotes in mm and inches) ÷1600 (63,00") 2450 (96,46&q

Page 4 - Cephalometric programs

Rotograph Evo D Product Data – Rev. 6 Page 2 of 11 Optional Digital Cephalometry Digital Ceph is available in two different versions: - Singl

Page 5 - Accessories

Rotograph Evo D Product Data – Rev. 6 Page 3 of 11 Patient Positioning Patient positioning is assured through multiple references • 2 temple c

Page 6 - Keyboard Functions

Rotograph Evo D Product Data – Rev. 6 Page 4 of 11 Exposure times Standard programs Projection Exposure time Panoramic 13.8 s adult / child Pa

Page 7 - Image Acquisition

Rotograph Evo D Product Data – Rev. 6 Page 5 of 11 Accessories • Standard chin support with bite stick (standard) • Reduced height chin suppor

Page 8 - Dental Studio software

Rotograph Evo D Product Data – Rev. 6 Page 6 of 11 Keyboard Functions LEGEND: Messages Display: indicates operative messages and warnings

Page 9 - Standards and Regulations

Rotograph Evo D Product Data – Rev. 6 Page 7 of 11 Digital Acquisition System PAN CCD Sensor • Technology: CCD sensor with Caesium Iodide (C

Page 10

Rotograph Evo D Product Data – Rev. 6 Page 8 of 11 Dental Studio software Rotograph Evo D is equipped with Dental Studio software for image acq

Page 11 - Via delle Azalee, 3

Rotograph Evo D Product Data – Rev. 6 Page 9 of 11 Mechanical features • Source to image distance: 500 mm (19,7”) for panoramic, TMJ and Sin